Journal of Physics E Scientific Instruments · 1985 · 48 citations · 10 references
EngineeringNuclear EngineeringAtomic PhysicsInstrumentationNuclear InstrumentationCharged Particle DetectionDetector SystemsDetector TechnologyHeavy-ion PhysicsRadiation DetectorsDetector DesignEnergy DependenceIon Detection EfficiencyMicrochannel Plate DetectorsMicrochannel PlatesSecondary Electron DetectorsSecondary Electron EmissionSecondary Electron Production
The performance of a double, chevron-mounted, microchannel plate ion detector of an atom-probe instrument has been studied using 3-25 keV He+, Ne+ and Ar+ ions. Pulse height distributions were recorded and the results indicate that the gain of the detector is strongly dependent on the ion energy. This energy dependence was found to be related to the secondary electron yield in the ion-channel-wall collision. However, the detection efficiency in atom-probe operation need not be affected by variations in mass and energy if an appropriate noise discrimination level is enforced and provided the ion energy exceeds approximately 5 keV.
10
Theory of Auger Ejection of Electrons from Metals by Ions
H. D. Hagstrum · Physical Review · 1954 · 1.3K citations
Imaging atom-probe mass spectroscopy
J. A. Panitz · Progress in Surface Science · 1978 · 113 citations