Journal of Microscopy · 1994 · 29 citations · 10 references
EngineeringMaterials ScienceBiomedical ImagingBiomedical OpticsSurface ScienceElectron MicroscopyTransmission Electron MicroscopyBright-field ImagingContrastImage Contrast MechanismsMaterial ContrastScanning Electron MicroscopyScanning Transmission Electron Microscopy
Summary The contrast thicknesses ( x k ) of thin carbon and platinum films have been measured in the transmission mode of a low‐voltage scanning electron microscope for apertures of 40 and 100 mrad and electron energies ( E ) between 1 and 30 keV. The measured values overlap with those previously measured for E (≥ 17keV) in a transmission electron microscope. Differences in the decrease of x k with decreasing E between carbon and platinum agree with Wentzel‐Kramer‐Brillouin calculations of the elastic cross‐sections. Knowing the value of x k allows the exponential decrease ∝ exp(— x / x k ) in transmission with increasing mass‐thickness ( x = ρt) of the specimen and the increasing gain of contrast for stained biological sections with decreasing electron energy to be calculated for brightfield and darkfield modes.