Applied Magnetic Resonance · 2012 · 15 citations · 16 references
EngineeringElectrical EngineeringInstrumentationNuclear InstrumentationSpectroscopyMagnetic Resonance SpectroscopyTiming AnalysisAdiabatic PulsesDirect Digital SynthesisInstrument Set AssemblyInstrument Set ConfigurationNuclear Magnetic Resonance AnalysisNuclear Magnetic Resonance SpectroscopyResonance MeasurementSynchronization
The design and construction of a high-performance, low-cost, and easy to assemble adiabatic extension set for homebuilt and commercial spectrometers is described. Described apparatus set was designed for the fast adiabatic passage generation and is based on direct digital synthesizer DDS. This solution gives generator high signal to noise ratio, phase stability even during frequency change which is only possible in expansive commercial high-end hardware. Critical synchronization and timing issues are considered and solutions are discussed. Different experimental conditions and techniques for the measurements are briefly discussed. The proposed system is very flexible and might be used for the measurement of low-frequency nuclear magnetic resonance.