Determination of cadmium and zinc isotope ratios in sheep's blood and organ tissue by electrothermal vaporization inductively coupled plasma mass spectrometry. Invited lecture

D. Conrad Grégoire, Julian Lee

Journal of Analytical Atomic Spectrometry · 1994 · 21 citations · 19 references

Concepts

Abstract

A method is described for the determination of Cd and Zn isotope ratios in sheep's blood and organ tissue. Samples were digested with nitric acid using a microwave oven. Cadmium and Zn were separated from matrix components using adsorption chromatography prior to isotope ratio measurement by electrothermal vaporization inductively coupled plasma mass spectrometry. A concentration factor of 35 was achieved. Limits of detection for the determination of Cd and Zn in blood were 0.34 and 0.40 pg g–1, respectively. Cadmium isotope ratios (111Cd: 106Cd; 111Cd: 110Cd) were determined with a precision of 2–3% for both peak height and area count measurements. Zinc isotope ratios (68Zn: 67Zn; 68Zn: 66Zn) were determined with a precision of 2% for peak height measurements and 1% for peak area count measurements.

References