International Journal of Environmental & Analytical Chemistry · 2005 · 20 citations · 5 references
EngineeringBiomedical EngineeringMechanical EngineeringActuatorsFluid DynamicsMicroelectromechanical SystemsMems ModelingMems SimulationEquivalent CircuitsFlow Rate DependenceMicropump SimulationMicropumpsPiezoelectric ElementsPumpingValvesVolumetric Flow Rate
This article describes a study concerning micropump design for medical purposes. In particular the project is focused on treatment of Hydrocephalus. An actuator glued on a membrane, a pumping chamber and a certain number of valves constitute the micropumps. The actuator is a piezoelectric disc, controlled according to data collected by means of a pressure sensor. We have studied two different structures of micropump: the first with membrane valves, and the second with diffuser/nozzle valves, without moving parts. Modelling both micropumps with electrical equivalent networks, we are able to estimate the pump behaviour, in terms of flow rate, with a simulator such as SPICE, and to optimize the micropump design for best performances.