Review of Scientific Instruments · 1999 · 85 citations · 15 references
Natural SciencesPhysicsNuclear InstrumentationPlasma PhysicsSpectroscopyHigh-temperature PlasmasMagnetized PlasmaNeutral Beam InjectionPlasma ConfinementPlasma DiagnosticsMagnetic DiagnosticsMagnetic Pitch AngleMotional Stark Effect DiagnosticsNew Instrument DesignPolarizationPolarization MeasurementsResponse Time
The motional Stark effect (MSE) diagnostic was developed to provide a measurement of the magnetic pitch angle or q profile in tokamaks. The technique relies upon polarization measurements of Stark broadened Dα emission to determine the pitch angle of the Lorentz vb×B electric field, where vb is the injected neutral beam particle velocity and B is the total magnetic field. However, in many advanced confinement regimes, large values of the plasma radial electric field, Er, are observed and can affect the interpretation of MSE measurements. Viewing fixed locations in the plasma from two different viewing angles allows one to separate the Er field from the vb×B field, thus providing simultaneous measurement of the Er and q profiles. To achieve this measurement, the DIII–D MSE diagnostic was recently upgraded from 16 to 35 channels with three independent viewing angles. The new instrument provides an Er resolution of 5–10 kV/m with a time response of 1 ms. Measurement results from the VH mode, reverse shear, and H mode plasmas are presented.
15
Improved Confinement with Reversed Magnetic Shear in TFTR
F. M. Levinton, M. C. Zarnstorff, S. H. Batha et al. · Physical Review Letters · 1995 · 704 citations
Equilibrium analysis of current profiles in tokamaks
L. L. Lao, J. R. Ferron, R. J. Groebner et al. · Nuclear Fusion · 1990 · 645 citations