Trace analysis of complex organic mixtures using capillary gas‐liquid chromatography and the dynamic solvent effect

Peter Apps, Victor Pretorius, Ken Lawson, Egmont R. Rohwer, M. R. Centner, Hermien Viljoen, G. A. Hulse

Journal of High Resolution Chromatography · 1987 · 25 citations · 7 references

Concepts

Abstract

Abstract A sampling system for high resolution gas‐liquid chromatography, based on the dynamic solvent effect, is described. Volatiles are accumulated off‐line in a concentrator/injector and delivered to the column using an on‐line inlet. Volatiles may be accumulated from gaseous or liquid matrices; they may be transferred to these by gas sparging or solvent extraction of any type of sample. The sampling technique is quantitatively precise; e.g. coefficients of variation of peak percentage areas better than 5% for a range of solutes at a concentration of 2:10 7 . Examples of the application of the sampling system are presented.

References

7