Heat conduction calorimeter for massively parallel high throughput measurements with picoliter sample volumes

E. B. Chancellor, John P. Wikswo, Franz Baudenbacher, M. Radparvar, D. P. Osterman

Applied Physics Letters · 2004 · 57 citations · 11 references

Concepts

Abstract

We have developed a bulk micromachined calorimeter with a sensitivity of 1.5nW∕Hz1∕2 and a 1ms time constant using a thin film thermopile as the sensing element. The thermopile consists of seven titanium and bismuth thermocouples with a total Seebeck coefficient of 574μV∕K. The device is capable of measuring enthalpies in chemical or biological reactions in volumes as small as a few picoliters. The device can be fabricated and operated in a massively parallel fashion in combination with ink-jet printing technologies in air and at room temperature, making it ideally suited for biological and biochemical experiments.