Publication | Closed Access
Thin film microcalorimeter for heat capacity measurements from 1.5 to 800 K
246
Citations
24
References
1994
Year
Thin Film PhysicsEngineeringDifferential Scanning CalorimetryMeasurementEducationHeat CapacityNew MicrocalorimeterElectronic DevicesCalibrationCalorimetryThin Film MicrocalorimeterThermal AnalysisThermodynamicsInstrumentationThin Film ProcessingMaterials ScienceHeat Capacity MeasurementsThin Film MaterialsThermal PhysicsCalorimetric MethodHeat TransferMicroelectronicsMicrofabricationMaterials CharacterizationApplied PhysicsTemperature MeasurementThin Film DevicesThin FilmsThermal SensorThermal Engineering
A new microcalorimeter for measuring heat capacity of thin films in the range 1.5–800 K is described. Semiconductor processing techniques are used to create a device with an amorphous silicon nitride membrane as the sample substrate, a Pt thin film resistor for temperatures greater than 40 K, and either a thin film amorphous Nb–Si or a novel boron-doped polycrystalline silicon thermometer for lower temperatures. The addenda of the device, including substrate, is 4×10−6 J/K at room temperature and 2×10−9 J/K at 4.3 K, approximately two orders of magnitude less than any existing calorimeter used for measuring thin films. The device is capable of measuring the heat capacity of thin film samples as small as a few micrograms.
| Year | Citations | |
|---|---|---|
Page 1
Page 1