Initial angle resolved measurements of fast neutrals using a multichannel linear AXUV detector system on LHD

E. A. Veshchev, T. Ozaki, P. R. Goncharov, S. Sudo

Review of Scientific Instruments · 2006 · 11 citations · 3 references

Concepts

Abstract

A new multichannel diagnostic for fast ion distribution studies has been developed and successfully tested on the Large Helical Device (LHD) in different plasma heating conditions. The diagnostic is based on a linear array AXUV detector consisting of 20 segments, charge sensitive preamplifiers, and a set of pulse height analysis channels. The main advantage of this system is the possibility to make time, energy, and angle-resolved measurements of charge exchange neutral particles in a single plasma discharge. This feature makes the new diagnostic a very helpful and powerful tool intended to contribute to the understanding of fast ion behavior in a complex helical plasma geometry like the one of LHD.

References

3