Experimental and Therapeutic Medicine · 2011 · 22 citations · 14 references
MedicineInfectious DiseasesMolecular BiologyMolecular DiagnosticsMolecular VirologyVirologyInfectious Disease Molecular TestingNucleic Acid Amplification TestingViral DiagnosticsViral HepatitisHepatitis Virus SerologyHepatitis Virus TestingPolymerase Chain Reaction Amplification EfficiencyReal-time Polymerase Chain Reaction (Pcr)Real-time Polymerase Chain Reaction AssaysSensitivityViral Load AssaysViral Load Assessment
In the present study, we standardized a TaqMan locked nucleic acid (LNA) real-time polymerase chain reaction (PCR) probe for the accurate quantification and detection of hepatitis B virus (HBV) DNA in serum (plasma), and evaluated its methodology. LNA probe technology had a much better detection performance in HBV DNA than the common TaqMan probe. The assay based on the LNA probe had a wider linear detection range, higher sensitivity, stability and amplification efficiency, and a lower concentration of probes than the TaqMan probe. Among the 15 cases with chronic hepatitis B surface antigen (HBsAg) (+) alone, only 4 cases that were detected by TaqMan real-time PCR were negative; however, the same samples were positive by LNA real-time PCR (p<0.05). A positive correlation between viral load measurements for the 35 samples with HBV-positive DNA was detected in both LNA and TaqMan real-time PCR.
14
Quantification of mRNA using real-time RT-PCR
Tania Nolan, Rebecca Hands, Stephen A. Bustin · Nature Protocols · 2006 · 2.3K citations