Transfer function of a crossed wire pendulum isolation system

Nobuyuki Kanda, M. A. Barton, Kazuaki Kuroda

Review of Scientific Instruments · 1994 · 18 citations · 6 references

DOIFull text

Open access

Concepts

Abstract

We measured the vibrational transfer function of a newly proposed crossed wire pendulum in the frequency range 0.03–3 Hz. We confirmed that it behaves as a much longer simple pendulum of the same period as far as one dimensional movement of its center of mass is concerned. We conclude that the pendulum shows promise as a component in an ultra-low frequency anti-vibration system.