A Josephson array voltage standard at 10 V

61

Citations

6

References

1987

Year

Abstract

The technology of Josephson voltage standards has been extended to an array of 14 184 junctions which is capable of generating over 150 000 quantized voltage levels spanning the range from -12 to +12 V. This makes possible the direct calibration of 10-V Zener reference standards without the use of a voltage divider.