Natural SciencesToxicologyAnalytical ChemistryElectroanalytical MethodsElemental AnalysisBioanalytical Method ValidationEnvironmental Analytical ChemistryInstrumental AnalysisMineral AnalysisSample PreparationAnalytical Detection Of Heavy MetalsCalcium DeterminationChromogenic DetectionLinear RangeMulticomponent AnalysisSample InjectionSpeciation In Environmental SamplesThroughput
The application of multicomponent techniques in sequential injection analysis is considered. Such a scarcely selective chromogenic reagent as 4-(2-pyridylazo)resorcinol allowed the simultaneous spectrophotometric determination of Ca and Mg in environmental samples. The process was carried out in a fully automated fashion by using a personal microcomputer running software for instrumental control and data acquisition and processing. The ensuing method features a linear determination range of 1–20 and 2–40 mg l–1 for magnesium and calcium, respectively, and a sample throughput of 60 h–1. The results are compared with those provided by batch and flow injection methodology.