Life SciencesBiochemistryAnalytical ChemistryEnzymologyMass Spectrometric AnalysisEnzyme AssaysEnzyme CharacterizationEnzyme KineticsProteolytic EnzymesEnzyme Assay KineticsEnzyme Kinetic AnalysisEnzyme Kinetics AssaysEnzyme Kinetics In DiagnosticsKinetic Data AnalysisKinetic Enzyme AssaysProtease Activity AssaysProteolysis Analysis
The concept of sequential injection analysis is described, and an application of this novel technique to an assay of traces of a proteolytic enzyme is demonstrated. The automated method is based on a stopped-flow procedure with fluorescence detection. Instead of using the conventional constant flow rate, a sinusoldal flow pattern, as generated by a robust cam-driven piston pump, has been applied. The validation of the novel flow injection system using a nonreactive fluorescent compound yielded highly reproducible results over a concentration range of 3 order of magnitude. The calibration graph for the enzyme activities in the range 5-75 micron mL-1 could be fitted a second-order polynomial. The detection limit is 2.5 micron mL-1, which corresponds to 7.2 ng mL-1 pure active enzyme.