IEEE Transactions on Control Systems Technology · 2007 · 377 citations · 34 references
EngineeringMechanical EngineeringMechatronics EngineeringComputational MechanicsInstrumentationMicroelectromechanical SystemsAtomic Force MicroscopyExperimental MechanicsMems DesignMems ModelingScanning Probe MicroscopyBandwidthHigh-speed Atomic Force MicroscopyHigh-speed Scanning Probe Microscopy
<para xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> A new mechanical scanner design for a high-speed atomic force microscope (AFM) is presented and discussed in terms of modeling and control. The positioning range of this scanner is 13 <formula formulatype="inline"><tex>$\mu$</tex> </formula>m in the <formula formulatype="inline"><tex>$X$</tex></formula>- and <formula formulatype="inline"><tex>$Y$</tex></formula>-directions and 4.3 <formula formulatype="inline"><tex>$\mu$</tex></formula>m in the vertical direction. The lowest resonance frequency of this scanner is above 22 kHz. This paper is focused on the vertical direction of the scanner, being the crucial axis of motion with the highest precision and bandwidth requirements for gentle imaging with the AFM. A second- and a fourth–order mathematical model of the scanner are derived that allow new insights into important design parameters. Proportional–integral (PI)-feedback control of the high-speed scanner is discussed and the performance of the new AFM is demonstrated by imaging a calibration grating and a biological sample at 8 frames/s. </para>
34
G. Binnig, C. F. Quate, Ch. Gerber · Physical Review Letters · 1986 · 14.4K citations · Full text
System identification—Theory for the user