Coherent artifact in modern pulse measurements

Justin Ratner, Günter Steinmeyer, Tsz Chun Wong, Randy A. Bartels, Rick Trebino

Optics Letters · 2012 · 91 citations · 10 references

DOIFull text

Open access

Concepts

Abstract

We simulate multishot intensity-and-phase measurements of unstable trains of complex ultrashort pulses using second-harmonic-generation (SHG) frequency-resolved optical gating (FROG) and spectral-phase interferometry for direct electric-field reconstruction (SPIDER). Both techniques fail to see the pulse structure. But FROG yields the correct average pulse duration and suggests the instability by exhibiting significant disagreement between measured and retrieved traces. SPIDER retrieves the correct average spectral phase but significantly underestimates the average pulse duration. In short, SPIDER measures only the coherent artifact. An analytical calculation confirms this last fact.

References

10