Biomedical Materials · 2009 · 43 citations · 34 references
Natural SciencesMaterials EngineeringMaterials ScienceBiocompatibilityBiomaterialsSurface ScienceBioceramicsMetal Oxide NanomaterialsMetal OxidesNanostructured BiomaterialsAnodizationNanostructured Surface FeaturesNanostructured Surface ModificationOxide Layer MorphologyTitanium BiocompatibilityTitanium OxidesTitanium Surface ModificationX-ray Diffraction Analysis
The surface of commercially pure titanium was modified by anodization treatment in a phosphoric acid solution at different voltages: 100 V, 200 V and 300 V. The surface characteristics of anodic TiO2 layers and their influence on the cell response were investigated. Micrographs by scanning electron microscopy revealed that the dense and uniform oxide layer obtained at 100 V exhibits a nanostructured surface which is similar to the surface of natural tooth cementum. In contrast, porous oxide layers without nanometer features were produced at higher voltages. Thin film x-ray diffraction analysis confirmed the existence of anatase in the oxide layer obtained at 300 V, but not in oxide layers obtained at 100 V and 200 V. The in vitro biocompatibility study of oxide layers demonstrated greater cell adhesion and proliferation of the oxide layer obtained at 100 V compared to the other two kinds of oxide layers.
34
Preparation of bioactive titanium metal via anodic oxidation treatment
Bangcheng Yang, Masaki Uchida, Hyun‐Min Kim et al. · Biomaterials · 2003 · 729 citations
Natural Sciences, Materials Engineering, Materials Science +13