Publication | Closed Access
Disassembly of a rotablator: Getting out of a trap
32
Citations
4
References
2003
Year
Rotablator BurrVisceral SurgerySurgeryManual TractionTrapped Burr
The rotablator burr rarely becomes trapped within calcified lesions. Manual traction can be ineffective and dangerous. We report a case that illustrates a novel technique involving use of a percutaneous snare in conjunction with partial disassembly of the rotablator device to remove a trapped burr without need for open surgical intervention.