Concepedia

Publication | Closed Access

Disassembly of a rotablator: Getting out of a trap

32

Citations

4

References

2003

Year

Abstract

The rotablator burr rarely becomes trapped within calcified lesions. Manual traction can be ineffective and dangerous. We report a case that illustrates a novel technique involving use of a percutaneous snare in conjunction with partial disassembly of the rotablator device to remove a trapped burr without need for open surgical intervention.