Publication | Closed Access
Design and testing of a white-light, steady-state diffuse reflectance spectrometer for determination of optical properties of highly scattering systems
177
Citations
24
References
1997
Year
Optical MaterialsEngineeringAbsorption SpectroscopyBiomedical EngineeringLight Scattering SpectroscopyOptical PropertiesDiffuse ReflectanceOptical SpectroscopyReflectanceBiophysicsMonte Carlo SimulationsRadiologyReflectance ModelingNear-infrared SpectroscopyBiophotonicsTissue-simulating PhantomsSpectroscopyBiomedical ImagingLight Scattering
We present a steady-state radially resolved diffuse reflectance spectrometer capable of measuring the absorption and transport scattering spectra of tissue-simulating phantoms over an adjustable 170-nm wavelength interval in the visible and near infrared. Measurements in a variety of phantoms are demonstrated over the relevant range of tissue optical properties, and the accuracy of the instrument is found to be approximately 10% in both scattering and absorption. Monte Carlo simulations designed to test the accuracy of the instrument are presented that support the experimental findings.