Applied Optics · 2000 · 39 citations · 16 references
EngineeringPhysicsInstrumentationMetrologyOpticsCalibrationInterferometryOptical AberrationsOptical InstrumentationOptical InstrumentsOptical MetrologyCalibration Of Optical InstrumentsInterferometric CalibrationOptical InterferometersOptical InterferometryPhase Shifting InterferometryPrecisionPrecision MetrologyRepeatability
A 300-mm-aperture digital phase-shifting Fizeau interferometer has been developed in-house for precision metrology of optical components fabricated by the optical workshop at Telecommunications and Industrial Physics, Commonwealth Scientific and Industrial Research Organization. We describe the procedures used in the calibration of the instrument. A reference data file representing the deviations from flatness of the reference surface is generated, measurement uncertainty estimated, and aberrations in the instrument assessed. Measurements on 250-mm-diameter uncoated optical surfaces have consistently shown short-term repeatability of 0.3-nm rms from measurement to measurement and allowed for absolute characterization of these surfaces to within a few nanometers.
16
C. J. Evans, Robert Kestner · Applied Optics · 1996 · 176 citations
Natural Sciences, Physics, Optics +13
Absolute Calibration Of An Optical Flat
Bernard S. Fritz · Optical Engineering · 1984 · 146 citations
Engineering, Physics, Metrology +14
Precise Measurement of Planeness
G. V. Schulz, Johannes Schwider · Applied Optics · 1967 · 145 citations
Engineering, Physics, Metrology +14