An improved method for the removal of ring artifacts in high resolution CT imaging
30
Citations
19
References
2012
Year
Computed TomographyImage ReconstructionStripe ArtifactsImage AnalysisImproved MethodMedical ImagingEngineeringReconstruction TechniqueHigh ResolutionBiomedical ImagingDetection AlgorithmCt ScanMedical Image ComputingRadiographic ImagingRadiologyHealth Sciences
Abstract In high resolution computed tomography (CT) using flat panel detectors, imperfect or defected detector elements cause stripe artifacts in sinogram which results in concentric ring artifacts in the image. Such ring artifacts obscure image details in the regions of interest of the image. In this article, novel techniques are proposed for the detection, classification, and correction of ring artifacts in the sinogram domain. The proposed method is suitable for multislice CT with parallel or fan beam geometry. It can also be employed for ring artifact removal in 3D cone beam volume CT by adopting a sinogram by sinogram processing technique. The detection algorithm is based on applying data driven thresholds on the mean curve and difference curve of the sinogram. The ring artifacts are classified into three types and a separate correction algorithm is used for each class. The performance of the proposed techniques is evaluated on a number of real micro-CT images. Experimental results corroborate that the proposed algorithm can remove ring artifacts from micro-CT images more effectively as compared to other recently reported techniques in the literature.