Metrologia · 2000 · 19 citations · 17 references
Natural SciencesElectrical EngineeringOpticsPhotonicsSpectroscopyDetector TechnologyPhotodetectorsRadiation InstrumentationSpectral CalibrationUltraviolet DetectorsUltraviolet InstrumentationDetector CharacterizationDetector Sensitivity CalibrationDetectorsPrecision Measurement TechniquesSpectral ResponsivitySystematic UncertaintiesUltraviolet PhotodetectorsWavelength Calibration
The Physikalisch-Technische Bundesanstalt (PTB) has established spectral responsivity standards for photodetectors in the wavelength range between 200 nm and 400 nm based on cryogenic electrical-substitution radiometers. In order to obtain standards over a continuous wavelength range with low uncertainty, combined use is made of both the dispersed synchrotron radiation of the storage ring BESSY I and quasi-monochromatic laser radiation. Taking advantage of the specific properties of the monochromatized synchrotron radiation and the laser radiation, relative standard uncertainties between 10−3 and 5 × 10−3 are achieved in the calibration of trap detectors in the spectral range 400 nm to 200 nm.
17
Hans Rabus, V. Persch, G. Ulm · Applied Optics · 1997 · 94 citations
Natural Sciences, Electrical Engineering, Radiation Detection +16