Journal of Biomaterials Applications · 2013 · 21 citations · 48 references
Natural SciencesBiomedical EngineeringMaterials ScienceBiocompatibilityBiomaterialsMechanical Behavior Of MaterialsScaffold DesignScaffold FabricationTextile MaterialsTissue Engineering ScaffoldsBiomaterial ScaffoldsFatigue BiomechanicsFatigue Endurance LimitFibersMedical TextilesStent BiomaterialsVascular Scaffolds
Stent-grafts externally fitted with a Z-shaped stents were compared to devices fitted with ringed stents in an in vitro oscillating fatigue machine at 200 cycles per minute and a pressure of 360 mmHg for scheduled durations of up to 1 week. The devices fitted with Z-stents showed a considerably lower endurance limit to buckling compared to the controls. The contact between the apexes of adjacent Z-stents resulted in significant damage to the textile scaffolds and polyester fibers due to the sharp angle of the Z-stents. The ringed stents did not cause any fraying in the textile scaffolds.
48
Endovascular versus Open Repair of Abdominal Aortic Aneurysm
Nora Toma · New England Journal of Medicine · 2010 · 1.4K citations · Full text
Peter L. Harris, S. Rao Vallabhaneni, Pascal Desgranges et al. · Journal of Vascular Surgery · 2000 · 754 citations · Full text
Medicine, Surgery, Vascular Surgery +14