Review of Scientific Instruments · 1995 · 25 citations · 0 references
Natural SciencesNuclear EngineeringPhysicsFusion EngineeringNuclear InstrumentationPlasma PhysicsFusion PlasmasHigh-temperature PlasmasLaser-plasma InteractionsPlasma ConfinementPlasma DiagnosticsPlasma HeatingElectron DensityHeating Of Tokamak PlasmasIon Cyclotron Resonance HeatingLaser-based DiagnosticsOhmic HeatingPlasma Density ProfilingPower OutputTokamak Physics
A multichannel far-infrared (FIR) hydrogen cyanide (HCN) laser interferometer was developed to measure plasma electron density profile on the HT-6M tokamak. The structure of the seven-channel FIR laser interferometer is described. The laser source used in the interferometer was a continuous-wave glow discharge HCN laser with a cavity length of 3.4 m and power output of about 100 mW at 337 μm. The detection sensitivity was 1/15 fringe with a temporal resolution of 0.1 ms. Experimental results were measured by the seven-channel FIR HCN laser interferometer with edge Ohmic heating, a pumping limiter, and ion cyclotron resonant heating on the HT-6M tokamak are reported.