Review of Scientific Instruments · 2006 · 82 citations · 9 references
PharmacyElectrical EngineeringAnalytical ChemistryInstrumentationMass Spectrometric AnalysisAccurate Mass MeasurementCharged Particle DetectionDetector SystemsMass AnalyzersMass SpectrometryCharge-to-mass SeparationFaraday CupsFaraday EffectIon Source DesignIon SourcesMass Spectrometry CharacterizationTime-of-flight Mass AnalyzersTime-of-flight Mass Spectrometry
We describe the design, electronics, and test results of a simple and low-cost time-of-flight ion charge-to-mass analyzer that is suitable for ion source characterization. The method selects a short-time sample of the beam whose charge-to-mass composition is then separated according to ion velocity and detected by a remote Faraday cup. The analyzer is a detachable device that has been used for rapid analysis of charge-to-mass composition of ion beams accelerated by voltages of up to about 100kV.
9
I.G. Brown · Review of Scientific Instruments · 1994 · 465 citations
Improved time-of-flight ion charge state diagnostic
I.G. Brown, James E. Galvin, R.A. MacGill et al. · Review of Scientific Instruments · 1987 · 138 citations
S. P. Bugaev, P. M. Schanin, G. Yu. Yushkov · Review of Scientific Instruments · 1994 · 57 citations