IEEE Transactions on Nuclear Science · 2013 · 28 citations · 15 references
Natural SciencesMaterials EngineeringMaterials ScienceElectronic MaterialsRadiation DetectionRadiation PhysicsDetector TechnologyGamma-ray DetectionGamma-ray DetectorsGamma-ray InstrumentationGamma-ray PolarimetryRadiation DetectorsFabricationGamma Detector SystemsGamma-ray Detector TechnologyHydrochloric Acid TreatmentLong-term StabilityRoom Temperature StabilityRoom Temperature StabilizationTemperature Stability
Thallium bromide (TlBr) is a material of interest for use in room temperature gamma ray detector applications due to is wide bandgap 2.7 eV and high average atomic number (Tl 81, Br 35). Researchers have achieved energy resolutions of 1.3% at 662 keV, demonstrating the potential of this material system. However, these detectors are known to polarize using conventional configurations, limiting their use. While high quality material is a critical starting point for excellent detector performance, we show that the room temperature stability of planar TlBr gamma spectrometers can be significantly enhanced by treatment with both hydrofluoric and hydrochloric acid. By incorporating F or Cl into the surface of TlBr, current instabilities are eliminated and the longer term current of the detectors remains unchanged. In addition the choice of electrode metal is shown to have a dramatic effect on the long term stability of TlBr detector performance <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">241</sup> Am spectra are also shown to be more stable for extended periods; detectors have been held at 4000 V/cm for 50 days with less than 10% degradation in peak centroid position.
15
Thallium bromide radiation detectors
K.S. Shah, J.C. Lund, F. Olschner et al. · IEEE Transactions on Nuclear Science · 1989 · 145 citations
Polarization Phenomena in TlBr Detectors
Keitaro Hitomi, Y. Kikuchi, Tadayoshi Shoji et al. · IEEE Transactions on Nuclear Science · 2009 · 78 citations
Engineering, Electrical Engineering, Electronic Materials +13
Thallium Bromide Nuclear Radiation Detector Development
A. Churilov, Guido Ciampi, Hadong Kim et al. · IEEE Transactions on Nuclear Science · 2009 · 65 citations