Unambiguous evaluation of a chirp measurement standard

Jörg Seewig, Matthias Eifler, Georg Wiora

Surface Topography Metrology and Properties · 2014 · 34 citations · 6 references

Concepts

Abstract

This article describes an automated evaluation method for the chirp standard. Chirp calibration standards provide a way to describe the transfer behavior of different spatial frequencies as they contain sinusoidal functions of varying wavelengths (Krüger-Sehm et al 2007 chirp calibration standards for surface measuring instruments Tech. Mess. tm 74 572–76; Pehnelt et al 2011 Comparative analysis of optical surface measuring systems with a chip calibration standard Tech. Mess. tm 78 457–62). By introducing a new, automated evaluation method, an improvement for the application of the chirp standard can be achieved. The data-preprocessing for topography and profile measurement data and the fit of the geometric elements are described. Automated evaluation can reduce the labor required to evaluate measured data and make it easier to compare different evaluations in the course of standardization. The algorithm can be used to characterize the so-called 'small scale fidelity' of an optical instrument. The term 'small scale fidelity' is currently discussed in the optical group of working group no. 16 of the ISO technical committee 213.