Analysis of calibration methods for direct emissivity measurements

Luis González-Fernández, R. B. Pérez‐Sáez, Leire del Campo, Manuel Tello

Applied Optics · 2010 · 37 citations · 17 references

Concepts

Abstract

We analyze two important points related to the experimental emissivity measurements in this paper: the radiometer calibration accuracy and its stability to determine the required frequency of calibration. The usual two-temperature calibration method is compared to a more accurate method, which uses the measurement of blackbody radiation at several temperatures. Additionally, the suitability of the two-temperature method is studied as a function of the gap between both temperatures. Differences higher than 200°C are needed to obtain an acceptable calibration. The temporal stability of the calibration and the influence of the environmental conditions are also analyzed.