Flow-Injection Amperometric Biosensor Based on Immobilized L-Lysine-α-oxidase for L-Lysine Determination

Aleksandr Simonian, I. E. Badalian, S. H. Khaduev

Analytical Letters · 1994 · 27 citations · 8 references

Concepts

Abstract

Abstract Amperometric flow-injection analytical system for quantitation of low L-lysine concentration was described. L-lysine-α-oxidase from Trichoderme sp. was immobilized by covalent linking on silica gel. The shift in oxygen concentration during enzymatic reaction was detected with a Clark membrane electrode. The rate of oxygen consumption was linearly dependent on L-lysine concentration over the 0,20 - 5,5 mM. Response time was 14 s, the total assay time about 2 min. The influences of pH, ionic strength and nature of the buffer on enzyme catalytic activity were tested. The analysis conditions were optimized. The immobilized L-lysine-α-oxidase retained the catalytic activity for about 5 months.