An enzyme electrode for specific determination of <scp>L</scp>‐lysine: A real‐time control sensor

35

Citations

14

References

1983

Year

Abstract

Lysine decarboxylase (L-lysine carboxylyase, E.C.4.1.1.18) is immobolized on a carbon dioxide gas-sensing electrode, by copolymerization with gelatin using the bifuncitional agent glutaraldehyde. The enzyme electrodes thus prepared are used in a continuous flow system to measure the concentration of L-lysine in a mixture of amino acids. The measuring time for each sample is about 3 min, including response and rinsing times. The electrode response is linear between 0.01-1 g/L and has a high specificity for L-lysine. The enzyme electrode response to lysine at concentrations below 0.5 g/L is stable on repeated use for at least 500 assays.