Rapid Communications in Mass Spectrometry · 2002 · 35 citations · 11 references
Natural SciencesBioinformaticsAnalytical ChemistryCombinatorial ChemistryMass Spectrometric AnalysisChromatographyHigh-throughput ScreeningInstrumental AnalysisLiquid ChromatographyMass SpectrometryCycle TimeLinear RangeLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method DevelopmentParallel SystemSystem Configuration
We report a high-throughput liquid chromatography/mass spectrometry (LC/MS) protocol for analyzing large combinatorial libraries using an eight-channel parallel LC/UV/MS (MUX-LCT) system. System configuration, linear response range in UV absorbance, LC column selection, and flow rate were optimized for 24 h/7 day unattended operations. Combinatorial libraries were analyzed on this system at a rate of 3200 compounds per day for a 3.5 min cycle time per injection. This parallel system is compared with a single-channel system in terms of performance and operation.
11
Martin K. Bayliss, David Little, David N. Mallett et al. · Rapid Communications in Mass Spectrometry · 2000 · 83 citations
Liyu Yang, Thierry Mann, David Little et al. · Analytical Chemistry · 2001 · 80 citations
David A Yurek, Derek L. Branch, Ming‐Shang Kuo · Journal of Combinatorial Chemistry · 2002 · 69 citations