Applied Optics · 1999 · 253 citations · 39 references
Natural SciencesBiomedical EngineeringChemistryAnalytical ChemistrySpectroscopic AnalysisSpectroscopyBioanalytical Method ValidationClinical AnalysisInfrared SpectroscopyNear-infrared SpectroscopyOptical SpectroscopyRaman SpectroscopyConcentrationMulticomponent AnalysisPhysiological MeasurementQuantification Of AnalytesRaman CalibrationRaman Data AnalysisRaman InstrumentationRaman Spectra
We demonstrate the use of Raman spectroscopy to measure the concentration of many important constituents (analytes) in serum and whole blood samples at physiological concentration in vitro across a multipatient data set. A near-infrared (830-nm) diode laser generates Raman spectra that contain superpositions of Raman signals from different analytes. Calibrations for glucose, cholesterol, urea, and other analytes are developed by use of partial least-squares cross validation. We predict six analytes in serum with significant accuracy in a 66-patient data set, using 60-s spectra. The calibrations are shown to be fairly robust against system drift over the span of seven weeks. In whole blood, a preliminary analysis yields accurate predictions of some of the same analytes and also hematocrit. The results hold promise for potential medical applications.
39
Partial least-squares regression: a tutorial
Paul Geladi, Bruce R. Kowalski · Analytica Chimica Acta · 1986 · 6.9K citations
Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose
William L. Clarke, Daniel J. Cox, Linda Gonder‐Frederick et al. · Diabetes Care · 1987 · 1.4K citations
Medicine, Endocrinology, Biosensors +16