Publication | Open Access
Design and performance of a programmable-temperature scanning tunneling microscope
49
Citations
16
References
1998
Year
EngineeringMicroscopyTunneling MicroscopyProgrammable-temperature ScanningMicroscopy MethodTunnelingInstrumentationPiezoelectric Scanner DesignSample Temperature RangeMaterials SciencePhysicsNanotechnologyMicroelectronicsMicrofabricationScanning Probe MicroscopySurface ScienceApplied PhysicsScanning Force MicroscopySample Temperature
In this article we introduce a novel scanning tunneling microscope (STM), which operates in a sample temperature range from 60 to at least 850 K. The most important new feature of this STM is that, while one selected part of the surface is kept within the microscope’s field of view, the sample temperature can be varied over a wide range of several hundreds of degrees during actual imaging. The extremely low drift of the scanner and sample was achieved by the combination of a thermal-drift compensated piezoelectric scanner design with a newly developed sample stage. The design of the sample stage defines a fixed center from which thermal expansions, in all three directions, are forced outwards. The performance of the microscope is demonstrated for several surfaces including Au(110), on which we follow one particular surface region over a temperature range of more than 270 K.