Review of Scientific Instruments · 1998 · 49 citations · 16 references
EngineeringMaterials EngineeringMaterials ScienceHeat TransferInstrumentationSurface ScienceAtomic Force MicroscopyScanning Probe MicroscopySurface MicroscopyTemperature MeasurementTemperature SensorsLow-temperature Scanning Probe MicroscopyMicroscope ImagingThermal Expansion Observation
In this article we introduce a novel scanning tunneling microscope (STM), which operates in a sample temperature range from 60 to at least 850 K. The most important new feature of this STM is that, while one selected part of the surface is kept within the microscope’s field of view, the sample temperature can be varied over a wide range of several hundreds of degrees during actual imaging. The extremely low drift of the scanner and sample was achieved by the combination of a thermal-drift compensated piezoelectric scanner design with a newly developed sample stage. The design of the sample stage defines a fixed center from which thermal expansions, in all three directions, are forced outwards. The performance of the microscope is demonstrated for several surfaces including Au(110), on which we follow one particular surface region over a temperature range of more than 270 K.
16
Diffusion of adsorbates on metal surfaces
R. Gomer · Reports on Progress in Physics · 1990 · 1.1K citations
Natural Sciences, Materials Engineering, Materials Science +17
M. Zinke-Allmang, L. C. Feldman, Marcia H. Grabow · Surface Science Reports · 1992 · 660 citations
J. G. Dash · Contemporary Physics · 1989 · 324 citations