Biomedical Optics Express · 2014 · 281 citations · 66 references
EngineeringBiophysicsPhysicsBiomedical OpticsBiophotonicsMedical PhysicsFlow MeasurementHemodynamicsLight-tissue InteractionOptical FlowOptical SpectroscopyScattering In TissueAutocorrelation FunctionAutocorrelation Function AnalysisBeer-lambert LawBeer-lambert Law In TissueBlood FlowBlood Flow MeasurementDiffuse Optical SpectroscopyFlow Monitoring
We develop and validate a Modified Beer-Lambert law for blood flow based on diffuse correlation spectroscopy (DCS) measurements. The new formulation enables blood flow monitoring from temporal intensity autocorrelation function data taken at single or multiple delay-times. Consequentially, the speed of the optical blood flow measurement can be substantially increased. The scheme facilitates blood flow monitoring of highly scattering tissues in geometries wherein light propagation is diffusive or non-diffusive, and it is particularly well-suited for utilization with pressure measurement paradigms that employ differential flow signals to reduce contributions of superficial tissues.
66
Diffuse optics for tissue monitoring and tomography
Turgut Durduran, Regine Choe, Wesley B. Baker et al. · Reports on Progress in Physics · 2010 · 1.2K citations
David J. Pine, David A. Weitz, P. M. Chaikin et al. · Physical Review Letters · 1988 · 1.2K citations