Physics in Medicine and Biology · 1998 · 10 citations · 12 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingDiagnostic Imaging SystemsImage OptimizationImage Quality AssessmentImage ReconstructionMedical ImagingCollimator DesignImage Distortion CorrectionImage Distortion ReductionOptimal Design
In this paper we present an analytical tool for the design and optimization of planar-concave collimators for SPECT. We conclude that a single general planar-concave collimator that eliminates the non-isotropic blurring for all SPECT applications does not exist. On the other hand, it is possible to achieve pseudo-optimal collimators for different clinical applications by a careful choice of the design parameters. By classifying the clinical applications into two groups, for instance body and brain studies, the non-isotropic blurring for most SPECT situations may be dramatically reduced by means of planar-concave collimators. The results based on Monte Carlo simulations show that the shape of the point source response function of the reconstructed image obtained from the planar-concave collimator is more isotropic than that obtained from a conventional parallel-hole collimator. Specifically, the ratio of the FWHM of the reconstructed point response function in the radial and tangential direction is increased from 0.5 for the parallel-hole to 0.85 for the planar-concave collimator for sources 200 mm away from the centre of rotation.