The capacitor insertion method to compensate for hysteresis and creep in piezoelectric actuators is applied to scanning tunneling microscopes. By inserting a capacitor in series with the piezoelectric actuator, hysteresis and creep can be greatly reduced or eliminated. After the principle of the method is reviewed, a structure of a scanning tunneling microscope used here is described. A calibration grid for scanning electron microscopes (2000 lines/mm) can be imaged with greatly reduced distortion by inserting capacitors in series with the piezoelectric actuators, which consist of an x-y-z scanner of the microscope. Furthermore the method can be used to reduce the sensitivity of the x-y-z movements of the scanner because an applied voltage is divided between an inserted capacitor and a piezoelectric actuator. Hence, using suitably small capacitors, atomic-scale images of Kish-graphite can be easily obtained with the same microscope.
5
Surface Studies by Scanning Tunneling Microscopy
G. Binnig, H. Rohrer, Ch. Gerber et al. · Physical Review Letters · 1982 · 4.6K citations · Full text
A study of graphite surface with stm and electronic structure calculations
Inder P. Batra, N. Garcı́a, H. Rohrer et al. · Surface Science · 1987 · 168 citations