Application of capacitor insertion method to scanning tunneling microscopes

Hiroshi Kaizuka

Review of Scientific Instruments · 1989 · 44 citations · 5 references

Concepts

Abstract

The capacitor insertion method to compensate for hysteresis and creep in piezoelectric actuators is applied to scanning tunneling microscopes. By inserting a capacitor in series with the piezoelectric actuator, hysteresis and creep can be greatly reduced or eliminated. After the principle of the method is reviewed, a structure of a scanning tunneling microscope used here is described. A calibration grid for scanning electron microscopes (2000 lines/mm) can be imaged with greatly reduced distortion by inserting capacitors in series with the piezoelectric actuators, which consist of an x-y-z scanner of the microscope. Furthermore the method can be used to reduce the sensitivity of the x-y-z movements of the scanner because an applied voltage is divided between an inserted capacitor and a piezoelectric actuator. Hence, using suitably small capacitors, atomic-scale images of Kish-graphite can be easily obtained with the same microscope.

References

5