Chinese Journal of Chemistry · 2007 · 11 citations · 24 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Emission SpectroscopyAtomic SpectroscopyHeavy Metal AnalysisInorganic Elemental AnalysisInstrumental AnalysisTrace Element AnalysisAtomic Emission SpectrometryChemometricsEmission SpectraInductively Coupled Plasma Atomic Emission SpectroscopyMultielement DeterminationPartial Least Squares RegressionPrincipal ComponentsUranium Analysis
Abstract Simultaneous determination of several elements (U, Ta, Mn, Zr and W) with inductively coupled plasma atomic emission spectrometry (ICP‐AES) in the presence of spectral interference was performed using chemometrics methods. True comparison between artificial neural network (ANN) and partial least squares regression (PLS) for simultaneous determination in different degrees of overlap was investigated. The emission spectra were recorded at uranium analytical line (263.553 nm) with a 0.06 nm spectral window by ICP‐AES. Principal component analysis was applied to data and scores on 5 dominant principal components were subjected to ANN. A 5‐5‐5 (input, hidden and output neurons) network was used with linear transfer function after both hidden and output layers. The PLS model was trained with five latent variables and 20 samples in calibration set. The relative errors of predictions (REP) in test set were 3.75% and 3.56% for ANN and PLS respectively.
24
Backpropagation neural networks
Barry J. Wythoff · Chemometrics and Intelligent Laboratory Systems · 1993 · 363 citations
Ali Reza Khanchi, Mohamad Mahani, Mohamad Ghanadi Maragheh et al. · Food Chemistry · 2006 · 125 citations