High temperatures and Raman scattering through pulsed spectroscopy and CCD detection

P. Šimon, B. Moulin, E. Buixaderas, N. Raimboux, Émilie Hérault, Bertrand Chazallon, H. Cattey, Nathalie Magneron, J. Oswalt, D. Hocrelle

Journal of Raman Spectroscopy · 2003 · 43 citations · 20 references

DOIFull text

Open access

Concepts

Abstract

Abstract A Raman scattering system devoted to high temperatures is presented. It is based on a pulsed system to remove thermal emission with retaining charge‐coupled device (CCD) detection. Two types of optical gating are used: a Pockels switch or an intensified CCD (ICCD), combining in this case the optical gate and the CCD detection. The performances of both systems are presented and their respective advantages are discussed and compared. The ICCD device allows recording of spectra in the 2000°C range, and typical results obtained on alumina and zirconia ceramics are shown. Copyright © 2003 John Wiley & Sons, Ltd.

References