2013 · 13 citations · 12 references
EngineeringBiomedical EngineeringElectrical EngineeringInstrumentationMicroelectronicsPhotonicsOptical SpectroscopySilicon PhotonicsTime MeasurementBicmos TechnologyChannelsFast LightHigh PerformancePhoton CountingSige Bicmos DesignTime ResolutionTime-correlated Spectroscopy
The analysis of extremely weak and fast light signals exploiting the measurement technique of Time-Correlated Single Photon Counting (TCSPC) is increasingly widespread in a large number of fields: from medicine and biology (fluorescence spectroscopy), to telemetry (laser ranging) and communication (quantum cryptography). The core of any TCSPC acquisition chain is the time measurement block: it has to be able to measure time intervals of few hundreds of nanoseconds with picosecond time resolution and very low differential non-linearity. In addition, collecting data from multiple channels simultaneously has recently become essential to employ the TCSPC technique in even more applications. To meet all these constraints we have designed a fully integrated 16-channel time-to-amplitude (TAC) array in 0.35μm Si-Ge BiCMOS technology showing very high performance. Each converter in the array provides a variable full-scale range (from 12.5 to 100ns) very good time resolution (lower than 27ps), extremely low differential non-linearity (better than 0.2% peak-to-peak of the LSB), low and constant power dissipation (less than 30mW) and high counting rate (11MHz).
12
Fluorescence lifetime imaging by time‐correlated single‐photon counting
W. Becker, Axel Bergmann, Mark A. Hink et al. · Microscopy Research and Technique · 2003 · 516 citations · Full text