Present Status of the NIMS Contract Beamline BL15XU at SPring-8

Shigenori Ueda, Yoshio Katsuya, Masahiko Tanaka, Hideki Yoshikawa, Yoshiyuki Yamashita, Satoshi Ishimaru, Yoshitaka Matsushita, Keisuke Kobayashi, R. Garrett, I. Gentle,

AIP conference proceedings · 2010 · 143 citations · 1 references

Concepts

Abstract

The revolver undulator beamline BL15XU at SPring‐8, which is the contract beamline of National Institute for Materials Science (NIMS), was established for materials science using soft‐and‐hard X‐ray photoelectron spectroscopy (XPS) and high‐resolution powder X‐ray diffraction (XRD). We have performed beamline reconstruction for further developments of the experiments in the research field of materials science. The flat double‐crystal monochromator (DCM) with liquid nitrogen cooling, X‐ray total reflection double‐mirror system with (+,−) geometry, and high‐energy‐resolution channel‐cut monochromator have been installed into the beamline. The refined beamline provides monochromatic X‐rays from 2 to 36 keV. The improvement of the photon flux density at the XRD and XPS experimental stations was confirmed. The photon flux was estimated to be several 1012 photons/sec with ΔE/E of ∼10−4.