Applied Optics · 2000 · 71 citations · 11 references
EngineeringPhysicsInstrumentationMetrologyOpticsInterferometryMeasurement System AnalysisOptical InstrumentationOptical MetrologyPhase AnalysisFringe AnalysisInterferometric CalibrationInterferometric InstrumentationInterferometric MeasurementInterferometric MetrologyNumerical ApertureOptical InterferometersOptical Interferometry
An interferogram produced by wide-aperture interferometers is studied both theoretically and experimentally. The fringe spacing is shown to increase nonlinearly with the numerical aperture and the fringe envelope to become narrower as the numerical aperture is increased. Phase measurements with wide-aperture interferometers therefore require calibration, and the phase can be measured only over a limited range. A calibration is given for accurate phase measurements, and the range over which the phase can be measured is calculated. Experimental measurements are presented and compared with theory.
11
G. S. Kino, Stanley S. C. Chim · Applied Optics · 1990 · 400 citations
Fringe spacing in interference microscopes
F R Tolmon, J. G. Wood · Journal of Scientific Instruments · 1956 · 65 citations
Natural Sciences, Physics, Metrology +13