Physics in Medicine and Biology · 1984 · 58 citations · 13 references
EngineeringNuclear EngineeringRadiation DetectionRadiation MeasurementSpectroscopyRadiation DetectorsRadiation SpectroscopyX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray SpectroscopyCadmium TellurideCadmium Telluride DetectorsCompton ScatteringPulse Height SpectraStrippingX-ray DiagnosticsX-ray SpectraX-ray Spectral Analysis
A method is introduced for the evaluation of x-ray spectra from x-ray machines operating in the range 50-100 kVp using a cadmium telluride (CdTe) detector with low detection efficiency. The pulse height distribution obtained with this kind of detector does not represent the true photon spectra owing to the presence of K-escape, Compton scattering, etc.; these effects were evaluated using a Monte Carlo method. A stripping procedure is described for implementation on a Univac 1100/82 computer. The validity of our method was finally tested by comparison with experimental results obtained with a Ge detector and with data from the literature; the results are in good agreement with published data.