Review of Scientific Instruments · 1976 · 43 citations · 20 references
EngineeringElectrical EngineeringMicroelectronicsMicrofabricationSensorsCharged Particle DetectionImage SensorsMicrofluidic DevicesMicrofluidicsRadiation DetectorsRadiation Imaging DetectorsCharge Collection ElectronicsDetector Readout ElectronicsDetector Spatial ResolutionMicrochannel Plate DetectorsMicrochannel PlatesPosition-sensitive DetectionPosition-sensitive ReadoutTime Resolution
A position-sensitive event-counting electronic readout system for microchannel plates (MCPs) is described that offers the advantages of high spatial resolution and fast time resolution. The technique relies upon a four-quadrant electron-collecting anode located behind the output face of the microchannel plate, so that the electron cloud from each detected event is partly intercepted by each of the four quadrants. The relative amounts of charge collected by each quadrant depend on event position, permitting each event to be localized with two ratio circuits. A prototype quadrant anode system for ion, electron, and extreme ultraviolet imaging is described. The spatial resolution achieved, ?10 μ, allows individual MCP channels to be distinguished.
20
Position Sensing by Charge Division
J. L. Alberi, V. Radeka · IEEE Transactions on Nuclear Science · 1976 · 115 citations
High-resolution imaging x-ray detector
E. Kellogg, Patrick M. Henry, S. S. Murray et al. · Review of Scientific Instruments · 1976 · 84 citations
Natural Sciences, Physics, Optics +16