Note: Switching crosstalk on and off in Kelvin probe force microscopy

Leo Polak, Sven de Man, R. J. Wijngaarden

Review of Scientific Instruments · 2014 · 24 citations · 10 references

DOIFull text

Open access

Concepts

Abstract

In Kelvin Probe Force Microscopy (KPFM) electronic crosstalk can occur between the excitation signal and probe deflection signal. Here, we demonstrate how a small modification to our commercial instrument enables us to literally switch the crosstalk on and off. We study in detail the effect of crosstalk on open-loop KPFM and compare with closed-loop KPFM. We measure the pure crosstalk signal and verify that we can correct for it in the data-processing required for open-loop KPFM. We also demonstrate that open-loop KPFM results are independent of the frequency and amplitude of the excitation signal, provided that the influence of crosstalk has been eliminated.

References

10