Review of Scientific Instruments · 2014 · 24 citations · 10 references
EngineeringOpen-loop KpfmPhysicsMicroscopyElectronic CrosstalkSpectroscopySurface ScienceApplied PhysicsBioelectronicsMicroscopy MethodTunneling MicroscopyScanning Force MicroscopyElectron MicroscopyScanning Probe MicroscopyInstrumentationExcitation SignalMedicineBiophysics
In Kelvin Probe Force Microscopy (KPFM) electronic crosstalk can occur between the excitation signal and probe deflection signal. Here, we demonstrate how a small modification to our commercial instrument enables us to literally switch the crosstalk on and off. We study in detail the effect of crosstalk on open-loop KPFM and compare with closed-loop KPFM. We measure the pure crosstalk signal and verify that we can correct for it in the data-processing required for open-loop KPFM. We also demonstrate that open-loop KPFM results are independent of the frequency and amplitude of the excitation signal, provided that the influence of crosstalk has been eliminated.
10
Accuracy and resolution limits of Kelvin probe force microscopy
Ulrich Zerweck, Christian Loppacher, Tobias Otto et al. · Physical Review B · 2005 · 426 citations
Naritaka Kobayashi, Hitoshi Asakawa, Takeshi Fukuma · Journal of Applied Physics · 2011 · 53 citations · Full text
Different Surface Charges, Engineering, Electrohydrodynamics +23