Publication | Closed Access
Locked Plating of Distal Femur Fractures Leads to Inconsistent and Asymmetric Callus Formation
289
Citations
44
References
2010
Year
Stabilization of distal femur fractures with periarticular locking plates can cause inconsistent and asymmetric formation of periosteal callus. A larger bridge span only minimally improves callus formation. The more flexible titanium plates enhanced callus formation compared with stainless steel plates.