Publication | Closed Access
Simultaneous recording of skin blood pulsations at different vascular depths by multiwavelength photoplethysmography
99
Citations
5
References
2007
Year
Medical UltrasoundEngineeringWearable TechnologyReflection PhotoplethysmographyBiomedical EngineeringMedical InstrumentationBiosignal ProcessingPatient MonitoringVascular ImagingSimultaneous RecordingBlood Flow MeasurementRadiologyHealth SciencesVascular ImageMedical ImagingBiophotonicsUltrasoundParallel RecordingSkin Blood PulsationsSpectroscopyPhysiologySkin SurfaceDifferent Vascular Depths
A new technique for parallel recording of reflection photoplethysmography (PPG) signals in a broad spectral band (violet to near-infrared) has been developed, and its potential for assessment of blood microcirculation at various depths from the skin surface is discussed. PPG signals have been simultaneously detected at cw laser wavelength sets comprising 405, 532, 645, 807, and 1064 nm. Various signal baseline responses to breath holding and different shapes of the PPG pulses originated from the same heartbeat but recorded at different wavelengths have been observed, indicating a depth variety of the skin blood pulsation dynamics.