Applied Spectroscopy · 1993 · 82 citations · 11 references
Natural SciencesBiomedical EngineeringOpticsPhotonicsSpectroscopyFiber OpticsHolographyOptical ImagingOptical InstrumentationOptical SpectroscopyRaman SpectroscopyDiffraction-limited ImagingDiffraction-limited ResolutionFiber Optic InstrumentsRaman ImagingRaman InstrumentationRaman SpectraRaman Spectroscopy ProbesVolume Holograms
An axial transmissive Raman spectrograph is described. The instrument employs volume-phase holograms as the laser line rejection filter and transmission grating. The on-axis configuration and the use of a high-spatial-frequency transmission grating allows nearly diffraction-limited imaging and pixel-limited resolution at f/1.8. The performance of the instrument is demonstrated for fiber-optic probe and microprobe Raman spectroscopy.