Journal of the American Chemical Society · 2005 · 680 citations · 14 references
Natural SciencesBioengineeringMaterials ScienceBiosensorsSpectroscopySurface ScienceNanosensorsOptical BiosensingOptical BiosensorsOptical SpectroscopyRaman SpectroscopySurface SpectroscopyBacillus SporesBacillus SubtilisCalcium DipicolinateDiffuse Reflectance SpectroscopyRaman Spectroscopy SensingRapid Detection AssaysShelf LifeSurface-enhanced Raman Spectroscopy
The study develops a rapid, battery‑powered, portable Raman spectrometer protocol for first‑responder detection of anthrax spores. The method uses silver film over nanosphere (AgFON) substrates optimized for 750 nm excitation, with calcium dipicolinate extracted by sonication in nitric acid, to detect Bacillus subtilis spores via surface‑enhanced Raman spectroscopy across a concentration range of 10⁻¹⁴–10⁻¹² M, yielding an adsorption constant of 1.7 × 10¹³ M⁻¹. The portable SERS sensor achieves an 11‑minute, 2.6 × 10³‑spore limit of detection, detects 10⁴ spores in 5 s on a one‑month‑old substrate, and remains functional for up to 40 days, demonstrating field‑ready speed and sensitivity below the anthrax infectious dose.
A rapid detection protocol suitable for use by first-responders to detect anthrax spores using a low-cost, battery-powered, portable Raman spectrometer has been developed. Bacillus subtilis spores, harmless simulants for Bacillus anthracis, were studied using surface-enhanced Raman spectroscopy (SERS) on silver film over nanosphere (AgFON) substrates. Calcium dipicolinate (CaDPA), a biomarker for bacillus spores, was efficiently extracted by sonication in nitric acid and rapidly detected by SERS. AgFON surfaces optimized for 750 nm laser excitation have been fabricated and characterized by UV−vis diffuse reflectance spectroscopy and SERS. The SERS signal from extracted CaDPA was measured over the spore concentration range of 10-14−10-12 M to determine the saturation binding capacity of the AgFON surface and to calculate the adsorption constant (Kspore = 1.7 × 1013 M-1). At present, an 11 min procedure is capable of achieving a limit of detection (LOD) of ∼2.6 × 103 spores, below the anthrax infectious dose of 104 spores. The data presented herein also demonstrate that the shelf life of prefabricated AgFON substrates can be as long as 40 days prior to use. Finally, these sensing capabilities have been successfully transitioned from a laboratory spectrometer to a field-portable instrument. Using this technology, 104 bacillus spores were detected with a 5 s data acquisition period on a 1 month old AgFON substrate. The speed and sensitivity of this SERS sensor indicate that this technology can be used as a viable option for the field analysis of potentially harmful environmental samples.
14
Chanda Ranjit Yonzon, Christy L. Haynes, Xiaoyu Zhang et al. · Analytical Chemistry · 2003 · 377 citations
Royston Goodacre, Beverley Shann, Richard J. Gilbert et al. · Analytical Chemistry · 1999 · 303 citations
Maritoni Litorja, Christy L. Haynes, Amanda J. Haes et al. · The Journal of Physical Chemistry B · 2001 · 128 citations
Natural Sciences, Materials Engineering, Materials Science +17