IEEE Transactions on Nuclear Science · 1975 · 28 citations · 8 references
EngineeringElectrical EngineeringInstrumentationRadiation DetectionSpectroscopyDetector SystemsDetector TechnologyPulse ProcessingRadiation DetectorsX-ray DetectionX-ray DetectorsCharge-sensitive AmplifiersDetector Energy ResolutionLive Time MeasurementRadiation Energy ResolutionSystem Energy ResolutionX-ray Spectroscopy Detectors
This paper describes the general features of a pulse processor for use with opto-electronic charge sensitive head amplifiers particularly with Si(Li) X-ray detectors. The instrument contains time-variant filters with direct coupled amplifiers and comprehensive pile-up protection and live time correction. The test results indicate that the system can cope with extremely wide range of signal processing times, counting rates and changes of spectral content without any appreciable degradation in energy resolution and with useful output rates for input rates up to well over 50,000 events/sec.
8
The use of a pileup rejector in quantitative pulse spectrometry
J. Bartošek, J. Mašek, F. Adams et al. · Nuclear Instruments and Methods · 1972 · 38 citations